Status
DECIDED
Stakeholders
Outcome

srdp > review by AoD, likely leave in and pass
non-srdp > leave in

In both cases include a qa note. Example:
The current calibrator models for resolved flux density calibrators in A-configuration in CASA are created
with narrow band data. At the high frequencies of a broad band observation, this can mean that the calibrator
model does not represent the structure of the calibrator very well, leading to amplitude errors that reduce
the dynamic range of the calibrated data. This will be remedied in a future version of CASA and the pipeline.
In the meantime, if high image fidelity is required we have found that recalibration using a model for a higher
frequency band scaled to the band of interest can work well.

Due date
Owner

Background

See qa note above.

Action items

  •