Outcome | srdp > review by AoD, likely leave in and pass non-srdp > leave in
In both cases include a qa note. Example: The current calibrator models for resolved flux density calibrators in A-configuration in CASA are created with narrow band data. At the high frequencies of a broad band observation, this can mean that the calibrator model does not represent the structure of the calibrator very well, leading to amplitude errors that reduce the dynamic range of the calibrated data. This will be remedied in a future version of CASA and the pipeline. In the meantime, if high image fidelity is required we have found that recalibration using a model for a higher frequency band scaled to the band of interest can work well. |
---|